Page:stress[:Cdefinition]:All:disable - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Example
Syntax
Parameter
Query response
Example
Syntax
Example
SCPI Commands

:PAGE:STRess[:CDEFinition]:ALL:DISable

:PAGE:STRess[:CDEFinition]:ALL:DISable
This command deletes the stress settings of all units (SMU,VSU,PGU,GNDU).
This command does not have query form.
:PAGE:STRess [ :CDEFinition] :ALL:DISable
OUTPUT @Hp4155;":PAGE:STR:ALL:DIS"
:PAGE:STRess[:CDEFinition]:COMMent
This command sets the USER DEFINED COMMENT for the stress group.
At *RST, this value is not defined.
:PAGE:STRess [ :CDEFinition] :COMMent comment
Parameter
Type
comment
string
comment <newline><^END>
OUTPUT @Hp4155;":PAGE:STR:COMM 'PulseStress'"
OUTPUT @Hp4155;":PAGE:STR:COMM?"
ENTER @Hp4155;A$
:PAGE:STRess[:CDEFinition]:GNDU:DISable
This command disables the stress settings of GNDU.
This command does not have query form.
:PAGE:STRess [ :CDEFinition] :GNDU:DISable
OUTPUT @Hp4155;":PAGE:STR:GNDU:DIS"
5-250
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
String of up to 58 characters.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4156c