Syntax
Parameter
Query response
Example

:PAGE:MEASure[:SWEep]:VAR1:SPACing

This command selects the sweep type of VAR1: linear staircase or logarithmic
staircase.
At *RST, this value is LINear.
:PAGE:MEASure [ :SWEep] :VAR1:SPACing LINear | L10 | L25 | L50
Parameter
Type
LINear
character
L10
character
L25
character
L50
character
LIN | L10 | L25 | L50 <newline><^END>
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:SPAC LIN"
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:SPAC?"
ENTER @Hp4155;A$
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:MEASure[:SWEep]:VAR1:SPACing
Explanation
The sweep is incremented (decremented) by the
stepsize until the stop value is reached.
The sweep is performed logarithmically (10 steps
per decade) between the stop and start values.
The sweep is performed logarithmically (25 steps
per decade) between the stop and start values.
The sweep is performed logarithmically (50 steps
per decade) between the stop and start values.