Syntax
Parameter
Query response
Example
:PAGE:CHANnels[:CDEFinition]:SMU<n>:MODE
This command sets the output MODE of SMU<n>. This command also has a query
form.
The value <n> is required to specify an SMU number. Valid SMU numbers are
SMU1 through SMU6, depending on the configuration.
At *RST, this value is set to:
SMU1
SMU2
SMU3 and SMU4
SMU5 and SMU6
:PAGE:CHANnels [ :CDEFinition] :SMU<n>:MODE V | I | VPULse | IPULse |
COMMon
:PAGE:CHANnels [ :CDEFinition] :SMU<n>:MODE?
Parameter
Type
V
character
I
character
VPULse
character
IPULse
character
COMMon
character
For the quasi-static CV measurements, V is available for the VAR1 channel, and V
and COMMON are available for the measurement channel.
V | I | VPUL | IPUL | COMM | DIS <newline><^END>
If an SMU is not used, the response data may be DIS(DISable).
OUTPUT @Hp4155;":PAGE:CHAN:SMU1:MODE V"
OUTPUT @Hp4155;":PAGE:CHAN:SMU1:MODE?"
ENTER @Hp4155;A$
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:CHANnels[:CDEFinition]:SMU<n>:MODE
SMU
OUTPUT MODE
COMMon
I
V
Not defined.
Explanation
Voltage output mode
Current output mode
Voltage pulse output mode (sweep mode only)
Current pulse output mode (sweep mode only)
Common
SCPI Commands
5-91