Page:channels[:Cdefinition]:Vsu<N>:Function - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
Query response
Example
:PAGE:CHANnels[:CDEFinition]:VSU<n>:FUNCtion
This command sets the function (FCTN) of VSU<n>.
<n> is required to specify VSU number. Valid VSU numbers are VSU1 through
VSU2.
At *RST, this value is CONStant.
:PAGE:CHANnels [ :CDEFinition] :VSU<n>:FUNCtion VAR1 | VAR2 | VARD |
CONStant
Parameter
Type
VAR1
character
VAR2
character
VARD
character
CONStant
character
VAR1 | VAR2 | VARD | CONS | DIS <newline><^END>
If VSU is not used, the response data may be DIS(DISable).
OUTPUT @Hp4155;":PAGE:CHAN:VSU:FUNC VAR1"
OUTPUT @Hp4155;":PAGE:CHAN:VSU:FUNC?"
ENTER @Hp4155;A$
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:CHANnels[:CDEFinition]:VSU<n>:FUNCtion
VAR1 function
VAR2 function
VAR1' function