:SYSTem:LANGuage

This command performs reset, then switches the remote command language from
the 4155/4156 (SCPI) to the 4145B syntax commands mode.
After switching the language to the 4145B syntax commands mode, only the *RST
command can switch the language back to SCPI mode.
This command does not have query form.
At *RST, the remote command language is SCPI.
:SYSTem:LANGuage COMPatibility
Parameter
Type
COMPatibility
character
OUTPUT @Hp4155;":SYST:LANG COMP"
:SYSTem:LFRequency
This command sets or queries the value that is set for the power line frequency
(reference frequency). *RST has no effect on the value of this parameter.
:SYSTem:LFRequency frequency
Parameter
Type
frequency
numeric
''HZ'' suffix can be used after frequency.
frequency <newline><^END>
frequency is NR3 response data.
OUTPUT @Hp4155;":SYST:LFR 50"
OUTPUT @Hp4155;":SYST:LFR 50HZ"