:PAGE:CHANnels[:CDEFinition]:VMU<n>:VNAMe
:PAGE:CHANnels[:CDEFinition]:VMU<n>:VNAMe
This command sets the VNAME of VMU<n>.
<n> is required to specify VMU number. Valid VMU numbers are VMU1 through
VMU2.
At *RST, this value is VMU<n>.
:PAGE:CHANnels [ :CDEFinition] :VMU<n>:VNAMe name
Parameter
Type
name
string
name <newline><^END>
name is the string response data, but does not contain double quote characters at the
beginning and end of the string.
OUTPUT @Hp4155;":PAGE:CHAN:VMU1:VNAME 'VD'"
OUTPUT @Hp4155;":PAGE:CHAN:VMU1:VNAME?"
ENTER @Hp4155;A$
:PAGE:CHANnels[:CDEFinition]:VSU<n>:DISable
This command deletes the settings of VSU<n>.
<n> is required to specify VSU number. Valid VSU numbers are VSU1 through
VSU2.
This command does not have query form.
:PAGE:CHANnels [ :CDEFinition] :VSU<n>:DISable
OUTPUT @Hp4155;":PAGE:CHAN:VSU1:DIS"
5-96
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
String of up to 6 alphanumeric characters.
1 st character must be alphabet.