Page:stress[:Cdefinition]:Vsu<N>:Disable - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Example
Syntax
Parameter
Query response
Example
:PAGE:STRess[:CDEFinition]:VSU<n>:DISable
This command deletes the stress settings of VSU<n>.
<n> is required to specify VSU number. Valid VSU numbers are VSU1 through
VSU2.
This command does not have query form.
:PAGE:STRess [ :CDEFinition] :VSU<n>:DISable
OUTPUT @Hp4155;":PAGE:STR:VSU1:DIS"
:PAGE:STRess[:CDEFinition]:VSU<n>:FUNCtion
This command sets the function (FCTN) of VSU<n> for stress.
<n> is required to specify VSU number. Valid VSU numbers are VSU1 through
VSU2.
At *RST, this value is NSYNC.
:PAGE:STRess [ :CDEFinition] :VSU<n>:FUNCtion SYNC | NSYNc
Parameter
Type
SYNC
character
NSYNc
character
SYNC | NSYN | DIS <newline><^END>
If VSU is not used, the response data may be DIS (DISable).
OUTPUT @Hp4155;":PAGE:STR:VSU1:FUNC SYNC"
OUTPUT @Hp4155;":PAGE:STR:VSU1:FUNC?"
ENTER @Hp4155;A$
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:STRess[:CDEFinition]:VSU<n>:DISable
Explanation
synchronous stress output
non-synchronous (non-stress output)
SCPI Commands
5-261

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4156c

Table of Contents