Page:measure[:Sweep]:Constant:smu<N>[:Source] - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
Query response
Example
SCPI Commands
:PAGE:MEASure[:SWEep]:CONStant:SMU<n>[:SOURce]
:PAGE:MEASure[:SWEep]:CONStant:SMU<n>[:SOU
Rce]
This command sets the constant SOURCE value of SMU<n> for the sweep
measurement.
<n> is required to specify SMU number. Valid SMU numbers are SMU1 through
SMU6, depending on the configuration.
You use this command only if the function of the specified SMU is CONStant and
the mode is not COMMon.
At *RST, this value is 0 V.
:PAGE:MEASure [ :SWEep] :CONStant:SMU<n>[ :SOURce] source | MINimum |
MAXimum
Parameter
Type
source
numeric
source <newline><^END>
source is NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:CONS:SMU1 10"
OUTPUT @Hp4155;":PAGE:MEAS:CONS:SMU1 MAX"
OUTPUT @Hp4155;":PAGE:MEAS:CONS:SMU1?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:CONS:SMU1? MAX"
ENTER @Hp4155;A
5-220
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
-200 to 200 V or -1 to 1 A. The range of this value
depends on the type of SMU.