:PAGE:MEASure:QSCV:SSTop

This command sets the stop condition for an abnormal status during Quasi-static CV
measurements. This command also has a query form.
The following abnormal status conditions can be detected:
1. Compliance on the non-measurement unit.
2. Compliance on the leakage current measurement unit.
3. Integration time too short at the capacitance measurement.
4. Overflow on ADC.
5. Oscillation on any unit.
At *RST, this value is OFF.
:PAGE:MEASure:QSCV:SSTop ABNormal | COMPliance | OFF
:PAGE:MEASure:QSCV:SSTop?
Parameter
Type
ABNormal
character
COMPliance
character
OFF
character
OFF | ABN | COMP <newline><^END>
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:SST ABN"
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:SST?"
ENTER @Hp4155;A$
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:MEASure:QSCV:SSTop
Explanation
Stops sweep if an abnormal condition is detected.
Stops sweep if the condition 2 or 3 shown above is
detected.
Continues sweep even if an abnormal condition is
detected.