Syntax
Example
Syntax
Parameter
Query response
Example
:PAGE:STRess[:CDEFinition]:PGU<n>:DISable
This command deletes the stress settings of PGU<n>.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
This command does not have query form.
:PAGE:STRess [ :CDEFinition] :PGU<n>:DISable
OUTPUT @Hp4155;":PAGE:STR:PGU1:DIS"
:PAGE:STRess[:CDEFinition]:PGU<n>:FUNCtion
This command sets the function (FCTN) of PGU<n> for stress.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
At *RST, this value is not defined.
:PAGE:STRess [ :CDEFinition] :PGU<n>:FUNCtion SYNC | NSYNc
Parameter
Type
SYNC
character
NSYNc
character
SYNC | NSYN | DIS <newline><^END>
If PGU is not used, the response data may be DIS (DISable).
OUTPUT @Hp4155;":PAGE:STR:PGU1:FUNC SYNC"
OUTPUT @Hp4155;":PAGE:STR:PGU1:FUNC?"
ENTER @Hp4155;A$
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:STRess[:CDEFinition]:PGU<n>:DISable
Explanation
synchronous stress output
non-synchronous (non-stress output)