Syntax
Common Commands

*TRG

*TRG
This command causes the following action depending on present display page.
If the present display page is the Stress Group page (STRESS CHANNEL
DEFINITION, STRESS SETUP or STRESS FORCE), this command starts the
stress force operation.
This action is as same as PAGE:SCONtrol:STRess[:STARt] command.
If the present display page is KNOB SWEEP page, this command generates
error "Init ignored"(-213).
If the present display page is not the Stress Group page or KNOB SWEEP page,
this command starts the single measurement operation.
This action is as same as PAGE:SCONtrol[:MEASurement]:SINGle command.
If this command is received while the 4155C/4156C is in the measurement or stress
state, this command generates error "Init ignored"(-213).
*TRG
4-16
Agilent 4155C/4156C SCPI Command Reference, Edition 1