Page:display:analysis:line(1|2):Point(1|2) :After:state - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Query response
Example
Syntax
Parameter
Query response
Example

:PAGE:DISPlay:ANALysis:LINE(1|2):POINt(1|2) :AFTer:STATe

var_name, expr <newline><^END>
var_name and expr are string response data, but do not contain double quote
characters at the beginning and end.
OUTPUT @Hp4155;":PAGE:DISP:ANAL:LINE1:POIN1:AFT 'DGM','M
AX(DGM)'"
OUTPUT @Hp4155;":PAGE:DISP:ANAL:LINE1:POIN1:AFT?"
ENTER @Hp4155;A$
:PAGE:DISPlay:ANALysis:LINE(1|2):POINt(1|2)
:AFTer:STATe
This command selects whether the start position of search is valid.
If AFTER var_name, expr is set, the STATe is automatically set to ON.
If POINts:MODE is XY (not OPLot), this parameter is ignored.
At *RST, this value is OFF.
:PAGE:DISPlay:ANALysis:LINE(1 | 2):POINt(1 | 2):AFTer:STATe OFF | ON | 0 | 1
Parameter
Type
OFF or 0
boolean
ON or 1
boolean
0 | 1 <newline><^END>
OUTPUT @Hp4155;":PAGE:DISP:ANAL:LINE1:POIN1:AFT:STAT ON"
OUTPUT @Hp4155;":PAGE:DISP:ANAL:LINE1:POIN1:AFT:STAT?"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
search entire area for the point
search for point only after the specified start position
SCPI Commands
5-113

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4156c

Table of Contents