Table 3-3
Operation Status Register
Bit
Definition
0
Calibrating
1 to 2
NOT USED
3
Sweeping
4
Measuring
5
Waiting for Trigger
6 to 7
NOT USED
8
Standby
9
Stressing
10 to 11
NOT USED
12
Printing
13
NOT USED
14
Program running
15
NOT USED
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Status Reporting Structure
Operation Status Register
Explanation
4155C/4156C is currently performing a
calibration.
Always zero.
A sweep is in progress.
4155C/4156C is actively measuring (sweep
or sampling measurement).
4155C/4156C is currently waiting for a
trigger input.
Always zero.
4155C/4156C is currently in the standby
state.
4155C/4156C is currently in the stress forcing
state.
Always zero.
A print/plot operation is in progress.
Always zero.
A user-defined program is currently in the run
state.
Always zero.
3-11