Page:channels[:Cdefinition]:Smu<N>:Function - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
Query response
Example
:PAGE:CHANnels[:CDEFinition]:SMU<n>:FUNCtion
This command sets the function (FCTN) of SMU<n>. This command also has a
query form.
The value <n> is required to specify the SMU number. Valid SMU numbers are
SMU1 through SMU6, depending on the configuration.
At *RST, this value is set to:
SMU1
SMU2
SMU3
SMU4
SMU5 and SMU6
:PAGE:CHANnels [ :CDEFinition] :SMU<n>:FUNCtion VAR1 | VAR2 | VARD |
CONStant
:PAGE:CHANnels [ :CDEFinition] :SMU<n>:FUNCtion?
Parameter
Type
VAR1
character
VAR2
character
VARD
character
CONStant
character
VAR1 | VAR2 | VARD | CONS | DIS <newline><^END>
If an SMU is not used, the response data may be DIS(DISable).
OUTPUT @Hp4155;":PAGE:CHAN:SMU1:FUNC VAR1"
OUTPUT @Hp4155;":PAGE:CHAN:SMU1:FUNC?"
ENTER @Hp4155;A$
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:CHANnels[:CDEFinition]:SMU<n>:FUNCtion
SMU
CONStant
VAR2
VAR1
CONStant
Not defined
VAR1 function (available for sweep and QSCV)
VAR2 function (sweep mode only)
VAR1' function (sweep mode only)
Constant
SCPI Commands
FUNCTION
Explanation
5-89