:PAGE:MEASure:SAMPling:CONStant:SMU<n>:COMPliance
:PAGE:MEASure:SAMPling:CONStant:SMU<n>:CO
MPliance
This command sets the constant COMPLIANCE value of SMU<n> for the sampling
measurement.
<n> is required to specify SMU number. Valid SMU numbers are SMU1 through
SMU6, depending on the configuration.
The mode of the specified SMU must be V or I.
:PAGE:MEASure:SAMPling:CONStant:SMU<n>:COMPliance compliance |
MINimum | MAXimum
Parameter
Type
compliance
numeric
compliance <newline><^END>
compliance is NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:CONS:SMU1:COMP 0.1"
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:CONS:SMU1:COMP MAX"
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:CONS:SMU1:COMP?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:CONS:SMU1:COMP? MAX"
ENTER @Hp4155;A
5-208
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
-200 to 200 V or -1 to 1 A. The range of this value
depends on the type of SMU.