Page:measure[:Sweep]:Vard:offset - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
Query response
Example
See also

:PAGE:MEASure[:SWEep]:VARD:OFFSet

This command sets the OFFSET value of VAR1'.
For each step of sweep, the output values of VAR1' are determined by the following
equation:
VAR1' = VAR1 ´ RATio + OFFSet
You use this command only if there is an SMU or VSU whose function (FCTN) is
VAR1'.
:PAGE:MEASure [ :SWEep] :VARD:OFFSet offset | MINimum | MAXimum
Parameter
Type
offset
numeric
offset <newline><^END>
offset is NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:VARD:OFFSET 0.5"
OUTPUT @Hp4155;":PAGE:MEAS:VARD:OFFSET?"
ENTER @Hp4155;A
:PAGE:MEASure[:SWEep]:VARD:RATio
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:MEASure[:SWEep]:VARD:OFFSet
Explanation
-400 to 400 V or -2 to 2 A. The range of this value
depends on the unit type of VAR1'.
SCPI Commands
5-237