Display:cmap:color<N>:Hsl - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
SCPI Commands
:DISPlay:CMAP:COLor<n>:HSL
:DISPlay:CMAP:COLor<n>:HSL
This command sets the instrument's color map based on the
Hue/Saturation/Luminance color model.
<n> is required to specify the color numbers. Valid COLor numbers are COLor1
through COLor15.
Color
Explanation
Number
1
Background
2
Softkey Background
3
Active Field Background
4
Graph Background
5
Foreground
6
Active Field Foreground
7
Title
8
Frame
:DISPlay:CMAP:COLor<n>:HSL hue | MINimum | MAXimum,
sat | MINimum | MAXimum, lum | MINimum | MAXimum
Parameter
Type
hue
numeric
sat
numeric
lum
numeric
5-26
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Color
Number
9
10
11
12
13
14
15
Explanation
Hue ranges circularly from 0 to 1 with a value of 0
resulting in the same hue as a value of 1. The
approximate color progression is (starting at 0): red,
orange, yellow, green, cyan, blue, magenta, and back
to red.
Saturation is the amount of pure color to be mixed
with white. The saturation value ranges from 0 to 1,
with 0 specifying no color (only white or gray,
depending on intensity) and 1 specifying no white.
Luminance specifies the brightness per unit area of
color. The luminance value ranges from 0 to 1. A
luminance of 0 results in black; a luminance of 1
results in the brightest color variable.
Explanation
Advisory
Y1 Axis
Y2 Axis
Marker/Cursor/Line
Active Mkr/Csr/Lne
Overlay
Grid

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4156c