Syntax
Query response
Example

:CALibration[:ALL]?

This command performs a full calibration of the instrument, then returns a
numeric_value that indicates the result of the calibration.
A zero is returned if calibration is completed successfully. Otherwise, one is
returned.
In addition to the error reported by this query response, the 4155C/4156C still
reports calibration errors through the normal status-reporting mechanism.
This command has query form only.
:CALibration [ :ALL] ?
result <newline><^END>
Result
0
1
The data type of result is NR1 response data.
OUTPUT @Hp4155;":CAL?"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
PASS
FAIL
SCPI Commands
:CALibration[:ALL]?
5-5