Device
Instrument
[1]
If the supplied power fluctuates beyond the rated voltage, a power line regulator may be required. High or low voltages can adversely
affect the electronic components of the instrument.
Wired and wireless network options and requirements
™
Thermo Fisher
Connect
connection
Network
connection
Wireless
A wireless adapter (also referred to as a dongle) is provided with the instrument. The wireless
connection conforms to 802.11 b/g/n wireless standards.
Wired
The instrument is factory-configured for IPv4 TCP/IP communication and uses an Ethernet
adapter (100/1,000 Mbps) with an RJ45-type connector for local area network (LAN)
connection.
• An active, tested network jack must be in place before the scheduled installation date.
• The assigned IT or network specialist from your organization must be available during the
Direct
No network is required.
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
Rated voltage
100–240 ±10% VAC
Local area network (LAN) intranet
installation to help connect the instrument to your network.
Wired and wireless network options and requirements
Maximum power
[1]
10 A (1,000 VA)
connection
LAN
Requirement
Appendix C Instrument specifications
Rated
frequency
50/60 Hz
Direct connection
C
Rated power
4.2 A (418 VA)
535