Wash Pump and Channels or Reactivate Instrument—Flush pump and channels
This step of the wizard flushes the pump and channels with polymer.
1. Close instrument doors.
2. Tap Flush pump and channels to flush the pump and channels with polymer.
Wait ~1 minute for the flush procedure to complete.
Wash Pump and Channels—Replace ABC
In this procedure, you replace the waste anode buffer container (ABC) with the previously installed ABC
or a new ABC.
1. Open instrument doors.
2. Remove the waste ABC.
3. Re-install the ABC that was removed earlier, or install a new ABC. See "Install the anode buffer
container (ABC)"
4. Tap Verify buffer to update the buffer information.
5. Tap Next to continue.

Replenish Polymer maintenance wizard

Replenish Polymer—Time and materials required
Use the Replenish Polymer maintenance wizard to install a new pouch of the currently installed
polymer type.
Note: To change polymer type, use the Change Polymer Type maintenance wizard.
Item
Time required
Materials required
Tap Next to continue.
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
on page
364.
~12 minutes
• New polymer pouch of the same type that is installed on the
instrument, equilibrated to room temperature (15–30°C)
• Purified water (distilled or deionized)
• Lint-free lab wipes
Chapter 13 Maintain the Instrument
Maintenance wizard procedures
Requirements
13
417