Thermo Scientific appliedbiosystems SeqStudio Flex Series User Manual page 442

Genetic analyzer with instrument software v1.0
Table of Contents

Advertisement

Chapter 14 Manage the software (instrument)
14
Connect the instrument to a computer or network (hardware connections)
4. Wait a few minutes, then check the IP addresses on the instrument and the computer. Both the IP
addresses must start with '169'.
• Instrument—In the home screen, tap
If the Direct Connection IP address field is blank or the IP address does not start with 169,
ensure that the cable is connected to the correct ports on the instrument and the computer.
If the cable is correctly connected and the IP addresses do not match, contact Technical
Support.
• Computer—See "Determine the IP address for a computer on a network"
5. Verify the connection by doing any of the following:
• Connect the instrument to a network drive (see "Connect the software to a network drive
(software settings)"
• If the Plate Manager software is installed on the computer, send a plate from Plate Manager to
the instrument (see "Send a plate file to the Inbox on the instrument"
• If the SAE Administrator Console is installed on the computer, enable SAE mode on
the instrument (see "Enable SAE mode on the instrument and specify the SAE server
(administrator only)"
6. If necessary, adjust the default network settings. See "Manage the network configuration
(administrator only)"
442
on page
443).
on page
240).
on page
464.
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
4Actions4Settings4About.
on page
443.
on page
126).

Advertisement

Table of Contents
loading

This manual is also suitable for:

A50369

Table of Contents