6. To export the log or report for the individual install run, tap Actions, then tap the item to export
(these options generate the same files you can generate at the end of an install run):
• Logs (ZIP)—Contains the install report (PDF), data files (FSA or AB1), install run file (CID), and
log files that can be used for troubleshooting.
Log file name format: InstrumentName_DyeSet_RunDate_Timestamp.zip, for example:
Instrument1_Z_BigDye Terminator v3.1_20201021_121952.zip
• Report (PDF)—
Contains the results of the install run for your records.
Report file name format: InstrumentName_DyeSet_RunDate_Timestamp.pdf, for
example: Instrument1_Fragment_20201021_121952.pdf
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
Chapter 12 Run calibrations and install checks
12
Perform an install run
353