1
Figure 8 CBC fill line
CBC fill line
1
IMPORTANT! Replace the buffer if the level is below the fill line.
Replenish consumables
If a consumable is expired or if the buffer fill level is below the fill line, replenish the consumables as
described in the following topics:
• "Replenish polymer or change polymer type (wizard)"
IMPORTANT! Wear gloves while handling polymer, the capillary array, septa, ABC, or CBC.
• "Install the anode buffer container (ABC)"
• "Install the cathode buffer container (CBC)"
• "Change and store a capillary array (wizards)"
Ensure proper installation of CBC septa
When you install the CBC buffer septa, press firmly to seat the septa.
IMPORTANT! Look at the CBC from the side and ensure that there is no gap between the container
and the lip of the septum.
For information about installing the septa, see "Install the cathode buffer container (CBC)"
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
on page 370
on page 364
on page 367
on page 371
Chapter 2 Get started
2
Prepare the instrument
on page
367.
71