10. Press the array RFID tag into place.
Figure 39 Press the RFID tag into place
RFID tag
1
11. Close the oven door.
12. Tap Verify array to update the array information.
13. Tap Next to continue.
Change Capillary Array—Prime pump
The Prime pump step in the maintenance wizard primes the pump with polymer.
1. Close the instrument doors.
2. Tap Prime pump.
3. Wait 3 minutes for the prime pump procedure to complete.
Change Capillary Array or Reactivate Instrument—Fill array and spatial calibration
1. Close the instrument doors.
2. Tap Fill array and spatial calibration.
3. Wait ~4 minutes for the fill array and spatial calibration procedures to complete.
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
Chapter 13 Maintain the Instrument
Maintenance wizard procedures
1
13
393