Install the cathode buffer container (CBC)

1. Check the expiration date on the CBC label to ensure that it is not expired and will not expire
during use.
2. Allow refrigerated CBC to equilibrate to room temperature (15–30°C).
3. In the home screen, tap
4. Follow the wizard instructions. See "Replenish Buffers maintenance wizard"
instructed to install the CBC, refer to this procedure.
5. Remove the installed CBC by gripping the center of the container and squeezing the clip at the
front, then lift the container from the instrument. Set aside for disposal.
IMPORTANT! If an autosampler error or a "power save" message is displayed, or if the CBC is
not in the forward, home position when you need to install the CBC, see "Check the position of the
cathode buffer container (CBC)"
6. Wipe away condensation on the new CBC exterior with a lint-free tissue. Condensation can cause
arcing and termination of the run.