Chapter 13 Maintain the Instrument
13
Maintenance wizard procedures
3. Remove the conditioning reagent pouch by pressing the polymer pouch lever down.
1
2
3
Figure 84 Location of pump, ABC, and pouch
Polymer pump block
1
Anode Buffer Container (ABC), remove and replace with an empty ABC to capture waste
2
Pouch, remove polymer and replace with conditioning reagent
3
1
2
Figure 85 Pouch slot and lever
Slot of the lever assembly
1
Label facing the instrument
2
Lever
3
4. Wipe the pouch connector on the instrument with a moistened lab wipe.
436
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide