Evaluate The Spatial Calibration Results - Thermo Scientific appliedbiosystems SeqStudio Flex Series User Manual

Genetic analyzer with instrument software v1.0
Table of Contents

Advertisement

Chapter 12 Run calibrations and install checks
12
Run a spatial calibration

Evaluate the spatial calibration results

See also "Spatial calibration troubleshooting"
1. Note the pass/fail status of the calibration at the top of the screen.
A Pass
Pass/fail result
1
Capillary CCD pixel marker
2
Tap for more results
3
2. Evaluate the spatial calibration profile. Ensure you see the following:
• One sharp peak for each capillary. Small shoulders are acceptable.
• One marker (+) at the apex of every peak. No off-apex markers.
• An even peak profile (all peaks about the same height).
306
status is required before you can run a spectral calibration.