Chapter 5 Create a plate file in the Plate Manager software
5
Set up a plate file

Enter plate properties

1. (Optional) In the Properties tab, edit the entries in the Plate file name, Barcode, or Owner fields.
2. In the Application type list, select a type: Sequencing, Fragment analysis, or Mixed (sequencing
and fragment analysis).
A mixed plate allows you to run sequencing and fragment analysis samples on the same plate.
3. In the Number of capillaries, Capillary length, and Polymer fields, specify the settings for the
configuration that the plate file will be used with.
4. Click Next.

Enter sample and injection information

This procedure describes using the plate view in the Plate tab. Alternatively, you can click
to enter information in table format. You can use keyboard shortcuts for text fields in both views
(for example, Sample name). You cannot use keyboard shortcuts for selection fields (for example,
Injection protocol). For information, see "Use the list view in the Plate tab"
shortcuts and examples"
1. In the Plate tab, select one or more injection groups: Click a well to select one injection group;
control+click or click-drag to select multiple injection groups; or click
select all injection groups.
Note: Injection group refers to each set of 8 or 24 wells on the plate, based on instrument
configuration.