Plate run status is set to running if no other plates are in the run queue
1
Time remaining for a running plate
2
Plate run status is set to In queue if other plates are in the run queue
3
Estimated end time for a plate that is in the run queue but has not started
4
Plate alerts and number of alerts
5
For more information on plate run statuses, see "Plate position status and plate run status"
page
39.
• Displays the End time in the Run Queue screen for each injection.
• Performs an auto-spectral calibration. For information, see "Auto-spectral calibration during an
injection"
on page
• Changes the status to Analyzing when data collection is complete and results are being analyzed.
Analysis can take up to 3 minutes. The next injection can start while the results for an injection are
being analyzed.
• Changes the status to Exporting if a save location other than the instrument is specified.
• Updates the Sample QC in the home screen and the Run Queue screen after each injection is
analyzed.