Run Module Settings - Thermo Scientific appliedbiosystems SeqStudio Flex Series User Manual

Genetic analyzer with instrument software v1.0
Table of Contents

Advertisement

2. Tap a row, tap Open, then tap Copy.
3. Enter a name for the run module, then tap Enter.
4. Tap sections of the screen and modify settings as needed (see "Run module settings"
page
273).
5. Tap Save.

Run module settings

Setting
(Lock)
Oven temperature (°C)
Injection voltage (volts)
Run voltage (volts)
Pre-run voltage (volts)
Injection time (sec)
Run time (sec)
Pre-Run time (sec)
Data delay (sec)
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
See "Lock or unlock a library entry"
Temperature setting for the instrument oven and the detection cell throughout the run.
Injection voltage setting for sample injection.
Final sample electrophoresis separation run voltage.
Pre run voltage setting before sample injection.
Sample injection time.
Length of time data is collected after voltage is ramped up to the run voltage and the
run starts.
Prerun voltage time.
Time from the start of separation to the start of sample data collection.
Chapter 11 Manage library resources on the instrument
Description
on page
255.
11
Run modules library
on
273