Installation; Unpacking Inspection; Requirements For Instrument Installation - Thermo Scientific Orion 8010cX User Manual

Ammonia analyzer
Table of Contents

Advertisement

Installation

Unpacking Inspection

Thermo Scientific Orion Ammonia Analyzers is assembled,
tested and packaged with great care.
Report any obvious damage of shipping container to carrier
and hold for inspection. The carrier (not Thermo Fisher
Scientific) is responsible for any damage incurred during
shipment. If any missing or damaged parts are found in these
components, please contact Thermo Fisher or its
representative office immediately.
1.Open the inner & outer packing, check the product and
accessories according to the packing checklist.
2.The shipping box contains the analyzer, LVD Safety Manual,
Quick Start Guide and User Manual USB drive.
3.Locate the dedicated external power switch or follow local
regulations for power supply.
4.Place the analyzer at a suitable location until installation is
complete.
5.Open the box and remove the accessories boxes from the
top, See Figure 4
Figure 4. Unpacking the 8010cX Ammonia
Analyzer
1. Outer Box
2. Inner Box
3. Accessories Box
4. EPE Packing Foam
Orion 8010cX Ammonia Analyzer
5. EPE Corner Foam
6. EPE Bottom Foam
7. Corrugated Protection Box
Requirements for
Instrument Installation
Basic Requirements for Instrument
Installation
Instructions for instrument handling:
CAUTION: The equipment exceeding 40 lbs.
should be lifted with the help of two persons
or follow any heavy lifting policies of your
employer has put in place.
• Safe lifting of the analyzer will require two people.
• The analyzer door should not be used during lifting as it will
cause the seal and hinge to become loose.
Selection of installation position:
Refer to the Specifications for the temperature and relative
humidity requirements of the installation environment.
It is suggested to install the instrument in an instrument
housing, a protective case, or an analysis room. The
temperature shall be controlled within 20 ±10°C and the
maximum humidity without condensation is less than 95%.
Orion 8010cX Ammonia Analyzer has IP65 protection grade
and can be installed outdoors. However, the installation
position of the instrument must be dry and avoid direct
sunlight, and the temperature variation between day and night
shall not be higher than 10°C / 24h.
Select the location as close as possible to the water sample
source to install the instrument, which can reduce the analysis
delay as much as possible.
The installation of the instrument shall comply with the
following requirements:
The instrument can be hung on the wall through the mounting
bracket, and can also be installed on a worktop or a vertical
cabinet for operation. No matter which installation method is
adopted, it is required to fasten the instrument with screws.
After being installed, the instrument must be secure to avoid
movement during operation.
Installation |
9