Create a new analysis settings

Note: The default sequencing analysis settings are optimized for PCR amplicon sequencing. For
sequencing of plasmid templates, create analysis settings with the following selections: for the End
base setting, deselect At PCR stop; for the Mixed base threshold setting, deselect Use Mixed base
identification.
1. In the home screen, tap
2. Tap a default analysis setting for your application, tap Open, then tap Copy.
3. Enter a name for the analysis settings, then tap Enter.
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
Chapter 11 Manage library resources on the instrument
Actions4Library4Analysis settings.
11
Analysis settings library
289