Rfid Troubleshooting - Thermo Scientific appliedbiosystems SeqStudio Flex Series User Manual

Genetic analyzer with instrument software v1.0
Table of Contents

Advertisement

This device complies with part 15 of FCC Rules and Innovation, Science and Economic Development
Canada's licence-exempt RSS(s). Operation is subject to the following two conditions: (1) this device
may not cause harmful interference, and (2) this device must accept any interference received, including
interference that may cause undesired operation.
Le présent appareil est conforme à la partie 15 des règles de la FCC et aux normes des
CNR d'Innovation, Sciences et Développement économique Canada applicables aux appareils radio
exempts de licence. L'exploitation est autorisée aux deux conditions suivantes : (1) l'appareil ne doit
pas produire de brouillage, et (2) l'appareil doit accepter tout brouillage subi, même si le brouillage est
susceptible d'en compromettre le fonctionnement.
FCC CAUTION
Changes or modifications not expressly approved by the party responsible for compliance could void
the user's authority to operate the equipment.
Note: This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection
against harmful interference when the equipment is operated in a commercial environment. This
equipment generates, uses, and can radiate radio frequency energy and, if not installed and used
in accordance with the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.

RFID troubleshooting

Table 23 RFID troubleshooting
Symptom
Unable to read RFID information.
"Failure to Read from RFID tag"
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
Appendix E Radio Frequency Identification (RFID) technology
Possible cause
Consumable is improperly
installed or label is defective.
Polymer/Conditioning reagent pouch
is not positioned properly.
Malfunctioning RFID label or reader.

RFID troubleshooting

Action
Ensure that the RFID label is not
visibly damaged and consumable
package is properly installed.
Ensure that label is close, and
parallel, to the instrument.
close the instrument door to update
the RFID information.
Install a new consumable (if
available).
If problem persists, contact Support.
Install a used CBC, ABC, pouch, or
array on the instrument:
• If the instrument can read the
RFID label, install a new CBC,
ABC, pouch, or array.
• If the instrument cannot read the
RFID label, contact Support.
E