3. Remove the capillary array:
a. Loosen the array port lock by turning it ~1/4 turn counter-clockwise until it stops. Adjust it
until the rectangular opening aligns with the rectangular head of the array tip.
Figure 27 Array port lock
Array port lock and array-port end of the capillary array
1
Retractable end of capillary array that protects the tip (slides to the right when you squeeze tab 1 and tab 2
2
toward each other)
Tab 1
3
Tab 2
4
b. Grasp the array-port end of the array and ease the tip out of the pump block.
c. Slide the retractor tab on the array to the right until it clicks into place in the retracted position.
Figure 28 Retracted end of the capillary array exposes the array tip
End retracted
1
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
1
2
3
1
Chapter 13 Maintain the Instrument
Maintenance wizard procedures
4
13
385