Install Or Change The Capillary Array (Wizards) - Thermo Scientific appliedbiosystems SeqStudio Flex Series User Manual

Genetic analyzer with instrument software v1.0
Table of Contents

Advertisement

Chapter 13 Maintain the Instrument
13
Change and store a capillary array (wizards)

Install or change the capillary array (wizards)

IMPORTANT! Before installing a capillary array, examine the loading-end header to ensure that the
capillary tips are not crushed or damaged.
For the time required, see "Maintenance wizard overview"
1. In the home screen, tap
2. In the Maintenance Wizards screen, tap Change capillary array.
IMPORTANT! Before installing a capillary array, ensure that the capillary tips are not crushed or
damaged. After installing the capillary array, ensure that the fastener on the detection cell heater
door is firmly tightened. If the fastener is not firmly tightened, peaks may be incorrectly detected.
3. Follow the instructions in the wizard to finish the procedure. See "Change Capillary Array
maintenance wizard"
IMPORTANT! If an autosampler error or a "power save" message is displayed, or if the CBC is
not in the forward, home position when you need to install the CBC, see "Check the position of the
cathode buffer container (CBC)"
374
Actions4Maintenance4Maintenance wizards.
on page
382.
on page
483.
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
on page
359.

Advertisement

Table of Contents
loading

This manual is also suitable for:

A50369

Table of Contents