Sequence analysis results
Parts of the sequencing electropherogram screen
1
2
3
4
5
6
Quality Value bars. For more information, see "Understanding Quality Values (QVs)"
1
Pure basecall with QV ≥20
•
Pure basecall with QV 15–19
•
Pure basecall with QV <15
•
Mixed basecall
•
Basecalls—The background of a mixed basecall is flagged in red. You can set the secondary peak threshold, as
2
a percentage of the primary peak, for consideration as a mixed basecall by the basecalling algorithm. Reaching
this threshold is a necessary but not sufficient condition for arriving at a mixed base determination. See "Analysis
settings—Sequencing"
Trace color hide/show—Tap to open, then tap a color to hide or show.
3
Analyzed trace.
4
Thumbnail trace—Drag the center of the pane in the thumbnail trace to display another trace area in the top pane.
5
Drag the right or left handle of the pane to zoom horizontally.
Sample name and result summary. See "Sequencing results and well details"
6
Trace tab—Tap to view the raw, EPT, or analyzed trace for the well. If the raw trace is blank, press-drag the thumbnail
7
pane over the region of data to view.
Zoom tools—Tap to open.
8
Well details—Tap to display well details. See "Sequencing results and well details"
9
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
on page
291.
Chapter 7 View results on the instrument
Sequence analysis results
7
8
9
on page
178.
on page
176.
on page
176.
7
175