Chapter 11 Manage library resources on the instrument
11
Plate files library
8. (Optional) Select a file name convention or a results group.
• A file name convention determines the naming of results files.
The default sequencing file name convention appends _WellID_SampleName_Timestamp
to the file name. The default fragment file name convention appends
_WellID_SampleName_SampleType_Timestamp to the file name.
• A results group determines the organization of results files.
The default results group organizes and groups files by start run time. On the instrument,
results are not grouped into subfolders specified by the results group, but you can view results
by results group name.
9. (Optional) Specify replicate injections. See "Specify replicate injections"
10. Tap the Sample name field, then proceed to "Enter sample names, sample types, and custom
fields"
on page
Define custom fields
Custom fields are text fields in which you can include additional sample attributes or identifiers. Custom
fields can be used in file name conventions or by some secondary analysis applications.
1. Open a plate file, then tap the Plate tab.
2. Tap to select one or more injection groups.
3. Tap the Injections pane.
262
93.
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
on page
92.