Chapter 13 Maintain the Instrument
13
Maintenance wizard procedures
3. Remove the polymer pouch by pressing the polymer pouch lever down.
1
2
Figure 63 Pouch slot and lever
Slot of the lever assembly
1
Label facing the instrument
2
Lever
3
4. Place a cap on the polymer, then store at 2–8°C.
5. Wipe the pouch connector on the instrument with a moistened lab wipe.
6. Install the conditioning reagent pouch:
a. Remove the seal from the conditioning reagent pouch.
b. With the label facing the rear of the instrument, insert the pouch fitting into the slot on the
pump lever mechanism.
c. Push the polymer pouch lever up.
7. Tap Verify pouch to to update the pouch information.
8. Remove the ABC and install an empty ABC to capture waste.
9. Close the instrument doors.
10. Tap Wash pump and channels.
IMPORTANT! After you tap Wash pump and channels you cannot cancel the operation. The
wash procedure takes ~45 minutes to complete.