Chapter 12 Run calibrations and install checks
12
Perform an install run
View, export, or delete an install run history
Note: Spatial and spectral calibration reports show the date of first installation. Install run reports show
the date of the most recent installation.
1. In the home screen, tap
2. (Optional) Tap
3. To export the list of calibrations that have been run on the instrument, tap Actions4Export
report.
The install run history report contains the results of all install runs performed on the instrument.
Report file name format:
InstrumentName_InstallHistoryReport_ReportGenerationDate_Timestamp.pdf, for example:
Instrument1_InstallHistoryReport_20201124_162247.pdf
4. To delete an install run history, tap a row, then tap
IMPORTANT! At least one history for each chemistry type must be retained. If you delete the only
run history for a chemistry type, it cannot be retrieved. You will have to perform a new install run for
the chemistry type before you can perform regular runs.
5. To view the details of an individual install run, tap a row, then tap View.
352
Actions4Maintenance4Install run4Install run history.
Filter to filter the list by polymer, result, and/or dye set.
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
Actions4Delete.