Thermo Scientific appliedbiosystems SeqStudio Flex Series User Manual page 79

Genetic analyzer with instrument software v1.0
Table of Contents

Advertisement

5. Verify that the holes of the plate retainer and the septum are aligned. If the holes are not aligned,
take the assembly apart, then re-assemble.
IMPORTANT! The array tips will be damaged if the plate retainer and septum holes are not
aligned.
8-tube strip assembly, standard or fast
IMPORTANT! Use the correct 8-tube standard or fast strips plate retainer and plate base with the
same catalog number. Using the wrong plate base may affect performance. See Appendix D, "Catalog
numbers" for plate assembly specifications and catalog numbers.
2
4
Figure 12 8-tube strip standard tube assembly
Plate retainer with one opening on side to allow
1
barcode reading
Septum strip
2
Retainer
3
8-tube strip standard tubes
4
96-well tray
5
Plate base
6
Note: Reuse of the septa strips is not recommended.
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
Chapter 3 Prepare the samples
Prepare and assemble sample plates
1
3
5
6
Figure 13 8-tube strip fast tube assembly
Plate retainer with one opening on side to allow
1
barcode reading
Septum strip
2
96-well tray
3
8-tube strip fast tubes
4
Plate base
5
3
1
2
3
4
5
79