Screen element
View plate button
Sample QC
Injection status
(Move)
Filter
Injection details
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
Active only when a single injection or plate with one injection is selected.
Displays the electropherogram or plate view for the selected injection.
• If you select a running or completed injection, the raw or analyzed plot is
displayed.
• If you select a plate with one injection, the plate view is displayed.
Lists the number of samples for each status:
Statuses are listed per plate and per injection. For more information, see "Sample
QC and quality alerts"
on page
(Clock) icon is displayed if the plate has been in the drawer for more than
24 hours. The software does not prevent you from running samples that have been on
the instrument for more than 24 hours.
Note: Samples that are prepared in Hi‑Di
Data quality may be reduced if samples exceed 24 hours on the instrument. Samples
that are prepared in aqueous solution have lower stability than samples that are
prepared in Hi‑Di
Formamide.
• In queue—Injection will start based on the order of the injection list.
• Running—Data collection is in process.
• Analyzing—Data collection is complete, analysis is in process.
• Exporting—Result files are being sent to the Save location that is specified in the
plate file.
• Completed—Analysis and export is complete, results are available.
• Aborting/aborted—An injection with status of Running was cancelled.
• Cancelled—An injection with status of In queue was cancelled.
• Paused—The instrument was paused (Actions4Pause instrument).
Press-drag to move an injection in the list and change the injection order.
Filter by injection status (not plate run status): All runs, Completed runs, Running and
In queue, Canceled runs.
Details for the selected injection. Injection parameters that have been edited for a
re-injection are displayed in orange.
Chapter 6 Start and monitor a run
Use the Run queue to monitor a run on the instrument
Description
Error,
172.
Formamide are stable for 16–24 hours.
Caution,
Good.
161