Thermo Scientific Orion Star A112 Reference Manual

Thermo Scientific Orion Star A112 Reference Manual

Benchtop and portable conductivity meters

Advertisement

Thermo Scientific Orion Star A112 Benchtop and
Star A122 Portable Conductivity Meters
Reference Guide

Advertisement

Table of Contents
loading

Summary of Contents for Thermo Scientific Orion Star A112

  • Page 1 Thermo Scientific Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters Reference Guide...
  • Page 2 Ross and the COIL trade dress are trademarks of Thermo Fisher Scientific, Inc. and its subsidiaries. AQUAfast, AQUASensors, BOD AutoEZ, ionplus, KNIpHE, LogR, No Cal, ORION, perpHect, PerpHecT, pHISA, pHuture, Pure Water, Sage, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS, and TURBO2 are registered trademarks of Thermo Fisher Scientific, Inc.
  • Page 3: Table Of Contents

    Thermo Scientific Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters Table of Contents Chapter 1 Introduction ............1 Chapter 2 Meter Overview.
  • Page 4 This page intentionally left blank.
  • Page 5: Chapter 1 Introduction

    Chapter 1 Introduction Thank you for your purchase of the Orion Star A112 benchtop conductivity or Star A122 portable conductivity meter. These meters are capable of measuring conductivity in μS/cm or mS/cm; TDS in mg/L; and temperature in °C or °F.
  • Page 6 This page intentionally left blank.
  • Page 7: Chapter 2 Meter Overview

    Connections 1. Power source: a. Power adapter (included with Orion Star A112 benchtop conductivity meters, sold separately for Star A122 portable conductivity meters) – Select the appropriate wall socket plug. Slide off the clear plastic cover, and slide on the plug plate into the groove on the back of the adapter.
  • Page 8: Display Information

    Primary display Larger, lower display showing measured value in selected mode. Instructions Located below the primary display. These phrases aid in the setup menu and calibration modes. Thermo Scientific Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters...
  • Page 9: Keypad Information

    Meter maintenance can be performed on a daily, weekly or monthly basis, as required by the environment in which the meter is operated. Immediately remove any spilled substance from the meter using the proper cleaning procedure for the type of spill. Thermo Scientific Orion Star A112 Benchtop and Star A122 Conductivity Portable Meters...
  • Page 10 This page intentionally left blank. Thermo Scientific Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters...
  • Page 11: Chapter 3 Meter Setup

    7.0 Factory Reset a. Press until the desired option is shown. b. Press to save your selection. 4. Press to exit the setup menu and return to measurement mode. Thermo Scientific Orion Star A112 Benchtop and Star A122 Conductivity Portable Meters...
  • Page 12 The default is no. Before selecting yes, please to the factory defaults and make sure any data that you would like to keep deletes all stored data has been recorded. (calibration and datalog). Thermo Scientific Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters...
  • Page 13: Setup Examples

    3. Press to enter the TDS factor. Note: Holding the button down will make the value change faster. 4. Press to save configuration and to return measurement mode. Thermo Scientific Orion Star A112 Benchtop and Star A122 Conductivity Portable Meters...
  • Page 14 This page intentionally left blank. Thermo Scientific Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters...
  • Page 15: Chapter 4 Calibration And Measurement

    4. Remove the conductivity probe from the sample, rinse with distilled water and blot dry. To continue taking measurements, place the probe into the next sample, stir gently and repeat step 3. 5. When finished measuring all samples, store probe according to the probe instructions. Thermo Scientific Orion Star A112 Benchtop and Star A122 Conductivity Portable Meters...
  • Page 16: Temperature Measurement And Calibration

    Temperature Measurement and Calibration The Orion Star A112 benchtop conductivity and Star A122 portable conductivity meter show temperature on the top, secondary display. To read only temperature and see the temperature on the primary, lower display, follow the instructions below.
  • Page 17: Chapter 5 Data Storage And Review

    Chapter 5 Data Storage and Review Orion Star A112 benchtop conductivity and Star A122 portable conductivity meters have a 50-point datalog memory. Automatic Datalog with AUTO-READ™ Mode 1. In measurement mode, press 2. Press three times until “4.0” is shown on the top line and “READ” is shown on the lower line. Press 3.
  • Page 18 This page intentionally left blank. Thermo Scientific Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters...
  • Page 19: Chapter 6 Customer Services

    If the existing memory’s data can be deleted: 1. Press change to “YES”. 2. Press to clear the datalog. Thermo Scientific Orion Star A112 Benchtop and Star A122 Conductivity Portable Meters...
  • Page 20: Troubleshooting Guide

    Thermo Scientific Orion products that may assist the detection of substances subject to the RoHS Directive are available at www.thermoscientific.com. Thermo Scientific Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters...
  • Page 21: Declaration Of Conformity

    Safety requirements for electrical equipment for measurement, CAN/CSA C22.2 No. 61010-1-04 control and laboratory use - general requirements Cheow Kwang Chan Place and Date of Issue: QA/Regulatory Manager June 15, 2011 Singapore Thermo Scientific Orion Star A112 Benchtop and Star A122 Conductivity Portable Meters...
  • Page 22: Meter Specifications

    TDS Factor Range Linear 0.01 to 10.00 Temperature Range -5 to 105°C Resolution Relative Accuracy ±0.1 Offset Calibration 1 point Memory 50 points Note: Specifications subject to change without notice. Thermo Scientific Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters...
  • Page 23: Ordering Information

    Orion Triode 3-in-1 pH/ATC Probe, Refillable, epoxy body 9107BNMD Orion Triode 3-in-1 pH/ATC Probe, Gel-filled, epoxy body 011050MD Orion 2-Electrode Conductivity Cell, K=1.0 083005MD Orion Polarographic DO probe , 1.5m cable Thermo Scientific Orion Star A112 Benchtop and Star A122 Conductivity Portable Meters...
  • Page 24 This page intentionally left blank. Thermo Scientific Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters...
  • Page 25: Chapter 7 Appendix

    Chapter 7 Appendix Temperature Compensation and Reference Temperature The Orion Star A112 benchtop conductivity and Star A122 portable conductivity meters have the ability to use a temperature compensation feature that calculates and displays the conductivity measurements at a reference temperature of 20 °C or 25 °C. The closer the sample temperature is to the selected reference temperature, the more accurate the conductivity measurement will be.
  • Page 26: Table Of Conductivity Standard Values Vs. Temperature

    1413 113.9 13.10 1441 115.9 13.35 1468 117.9 13.59 1496 120.0 13.84 1524 122.0 14.09 1552 124.1 14.34 1580 126.2 14.59 1608 128.3 14.85 1636 130.4 15.10 1665 Thermo Scientific Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters...
  • Page 27 1866 147.6 17.15 1896 149.8 17.40 1925 152.0 17.66 1954 154.2 17.92 1983 156.4 18.18 2013 158.7 18.44 2042 160.9 18.70 2071 163.2 18.96 2101 165.4 19.22 2130 Thermo Scientific Orion Star A112 Benchtop and Star A122 Conductivity Portable Meters...
  • Page 28: Notes

    Notes Thermo Scientific Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters...
  • Page 30 Water Analysis Instruments North America 9 0 0 1 : 2 0 0 8 166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 info.water@thermo.com Netherlands Tel: (31) 033-2463887 info.water.uk@thermo.com India Tel: (91) 22-4157-8800 wai.asia@thermofisher.com Japan Tel: (81) 045-453-9175 wai.asia@thermofisher.com China Tel: (86) 21-68654588...

This manual is also suitable for: