Table 11 Summary tab—sequencing install run (continued)
Result
Spectral calibration
Spectral calibration status Pass or fail.
Spectral calibration
passing criteria
# caps passed/# caps
borrowed
Capillaries tab—sequencing install run
Table 12 Capillaries tab—sequencing install run
Result
QV20 CRL (range)
QV20 basecall
(This result is referred to
as QV20 on the Details
screen.)
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
The number of capillaries that can be borrowed: 2 for 24-capillary instruments and
1 for 8-capillary instruments. For more information, see "Spectral calibration sharing
between capillaries"
on page 321
on page
322.
The number of capillaries that met the spectral QV and condition number limits. See
"Spectral QV (quality value) and Condition number definitions and limits"
The number of capillaries that were borrowed.
The QV20 CRL result and the base number range in which the QV20 CRL was
determined.
The percent accuracy of the basecalls in the sample sequence within the QV20
CRL range relative to the known reference sequence of the sequencing install
standard.
Chapter 12 Run calibrations and install checks
Description
and "Spectral calibration with borrowing enabled"
Description
12
Perform an install run
on page
320.