Thermo Scientific appliedbiosystems SeqStudio Flex Series User Manual page 477

Genetic analyzer with instrument software v1.0
Table of Contents

Advertisement

3. Tap a well, then tap View.
By default, the Analyzed tab is selected and the size standard peaks are shown. If the size
standard peaks are not visible, you can tap
dyes except the size standard dye (red or orange). You can also move the thumbnail window to the
size standard peak region.
4. Ensure that all expected peaks are detected.
Note: To determine the size standard that was used, tap
expected peaks, view the size standard in the size standard library. See "Size standards library"
page
275.
5. As needed, tap
Example of a size standard electropherogram
The results show
Although no alerts were triggered, the low SQ should be investigated. In this example, the low SQ is
caused by the issues with peak resolution later in the electropherogram.
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
on the right border of the trace to zoom.
Sample QC status, SQ 0.86.
Appendix A Troubleshooting
Inspect the data—Troubleshooting procedures
on the left border of the trace, then deselect all
(Information). To determine the
A
on
477