Chapter 13 Maintain the Instrument
13
Maintenance wizard procedures
Change Capillary Array or Shut Down Instrument—Discard capillary array
WARNING! SHARP The electrodes on the load-end of the capillary array have small, blunt ends that
can lead to piercing injury.
1. Discard the capillary array according to your laboratory protocol.
2. Tap Next to continue.
Change Capillary Array—Install capillary array
WARNING! SHARP The electrodes on the load-end of the capillary array have small, blunt ends that
can lead to piercing injury.
In this procedure, you install a capillary array in the instrument.
1. Clean the bottom surfaces of the detection cell heater with a moistened, lint-free lab wipe.
1
2
2
3
Figure 32
Heat dissipation sheet
1
Bottom surfaces
2
Detection cell heater door (open)
3