Thermo Scientific Orion Star A111 Reference Manual

Thermo Scientific Orion Star A111 Reference Manual

Benchtop and portable ph meters

Advertisement

Thermo Scientific Orion Star A111 Benchtop and
Star A121 Portable pH Meters
Reference Guide

Advertisement

Table of Contents
loading

Summary of Contents for Thermo Scientific Orion Star A111

  • Page 1 Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters Reference Guide...
  • Page 2 Ross and the COIL trade dress are trademarks of Thermo Fisher Scientific, Inc. and its subsidiaries. AQUAfast, AQUASensors, BOD AutoEZ, ionplus, KNIpHE, LogR, No Cal, ORION, perpHect, PerpHecT, pHISA, pHuture, Pure Water, Sage, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS, and TURBO2 are registered trademarks of Thermo Fisher Scientific, Inc.
  • Page 3: Table Of Contents

    Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters Table of Contents Chapter 1 Introduction ............1 Chapter 2 Meter Overview .
  • Page 4 This page intentionally left blank.
  • Page 5: Chapter 1 Introduction

    Chapter 1 Introduction Thank you for your purchase of the Orion Star A111 benchtop pH or Star A121 portable pH meter. These meters are capable of measure pH, raw millivolts (mV), relative millivolts (Rmv), oxidation-reduction potential (ORP) and temperature. The Orion Star A111 benchtop pH meters are IP54-rated. The Orion Star A121 portable pH meters feature a waterproof, IP67-rating.
  • Page 6 This page intentionally left blank.
  • Page 7: Chapter 2 Meter Overview

    Connections 1. Power source: a. Power adapter (included with Orion Star A111 benchtop pH meters, sold separately for Star A121 portable pH meters) – Select the appropriate wall socket plug. Slide off the clear plastic cover, and slide on the plug plate into the groove on the back of the adapter.
  • Page 8: Display Information

    Primary display Larger, lower display shows measured value in selected mode. Instructions Located below the primary display. These phrases aid in the setup menu and calibration modes. Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters...
  • Page 9: Keypad Information

    Meter maintenance can be performed on a daily, weekly or monthly basis, as required by the environment in which the meter is operated. Immediately remove any spilled substance from the meter using the proper cleaning procedure for the type of spill. Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters...
  • Page 10 This page intentionally left blank. Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters...
  • Page 11: Chapter 3 Meter Setup

    7.0 Factory Reset a. Press until the desired option is shown. b. Press to save your selection. 4. Press to exit the setup menu and return to measurement mode. Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters...
  • Page 12 This is the slope data from the last calibration. SLP.2, SLP.2 and SLP.A (average slope) will only show SLP.A if 3 points are calibrated. Press to scroll between them. Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters...
  • Page 13 The default is no. Before selecting yes, please make settings to the sure any data that you would like to keep has been factory defaults and recorded. deletes all stored data (calibration and datalog). Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters...
  • Page 14: Setup Examples

    If a 3-point calibration was done, press to view the second slope segment, and again to display the average slope (SLP.A). 3. Press to return to measurement mode. Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters...
  • Page 15: Chapter 4 Calibration And Measurement

    , place electrode (and ATC probe, if separate) into the next sample and repeat steps 3 and 4. 5. When finished measuring all samples, store electrode according to electrode instructions. Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters...
  • Page 16: Mv Measurement And Rmv & Orp Calibration And Measurement

    3. Remove the electrode (and ATC probe, if separate) from the sample, rinse with distilled water, blot dry. To continue taking measurements repeat steps 2 through 4. 4. When finished measuring all samples, store electrode according to electrode instructions. Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters...
  • Page 17: Temperature Measurement And Calibration

    Note: The calculated offset will be applied to all future temperature readings. To abort, press to end without saving and return to the measurement mode. 4. When finished, press to save and end calibration. Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters...
  • Page 18 This page intentionally left blank. Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters...
  • Page 19: Chapter 5 Data Storage And Review

    Chapter 5 Data Storage and Review Orion Star A111 benchtop pH and Star A121 portable pH meters have a 50 point datalog. Manual Datalog with Continuous Read Mode 1. In measurement mode, press 2. Press three times until “4.0” is shown on the top line and “READ” is shown on the lower line. Press 3.
  • Page 20: Viewing The Data From The Last Calibration

    Viewing Stored Readings (the Data Log) 1. In measurement mode, press 2. Press to scroll through the memory points. 3. Press to review the reading stored at that point. Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters...
  • Page 21: Chapter 6 Customer Services

    If the existing memory’s data can be deleted: 1. Press change to “YES”. 2. Press to clear the datalog. Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters...
  • Page 22: Troubleshooting Guide

    Thermo Scientific Orion products that may assist the detection of substances subject to the RoHS Directive are available at www.thermoscientific.com. Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters...
  • Page 23: Declaration Of Conformity

    Safety requirements for electrical equipment for measurement, CAN/CSA C22.2 No. 61010-1-04 control and laboratory use - general requirements Cheow Kwang Chan Place and Date of Issue: QA/Regulatory Manager June 15, 2011 Singapore Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters...
  • Page 24: Meter Specifications

    RmV: ±1999.9 mV Resolution Relative Accuracy ±0.2 mV or ±0.05 % of reading whichever is greater Temperature Range -5 to 105°C Resolution Relative Accuracy ±0.1 Offset Calibration 1 point Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters...
  • Page 25: Ordering Information

    Orion Triode 3-in-1 pH/ATC Probe, Refillable, epoxy body 9107BNMD Orion Triode 3-in-1 pH/ATC Probe, Gel-filled, epoxy body 011050MD Orion 2-Electrode Conductivity Cell, K=1.0 083005MD Orion Polarographic DO probe , 1.5m cable Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters...
  • Page 26 This page intentionally left blank. Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters...
  • Page 27: Chapter 7 Appendix

    Automatic pH Buffer Recognition Feature The Orion Star A111 benchtop pH and Star A121 portable pH meters are capable of automatically recognizing pH 1.68, 4.01, 7.00, 10.01 and 12.46 buffers or pH 1.68, 4.01, 6.86, and 9.18 buffers during a pH calibration, depending on the pH buffer set that is selected in the setup menu.
  • Page 28: Notes

    Notes Thermo Scientific Orion Star A111 Benchtop and Star A121 Portable pH Meters...
  • Page 29 This page intentionally left blank.
  • Page 30 Water Analysis Instruments North America 9 0 0 1 : 2 0 0 8 166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 info.water@thermo.com Netherlands Tel: (31) 033-2463887 info.water.uk@thermo.com India Tel: (91) 22-4157-8800 wai.asia@thermofisher.com Japan Tel: (81) 045-453-9175 wai.asia@thermofisher.com China Tel: (86) 21-68654588...

This manual is also suitable for:

Star a121

Table of Contents