Thermo Scientific appliedbiosystems SeqStudio Flex Series User Manual page 437

Genetic analyzer with instrument software v1.0
Table of Contents

Advertisement

5. Install the new polymer pouch:
a. Remove the seal from the conditioning reagent pouch.
b. With the label facing the rear of the instrument, insert the pouch fitting into the slot on the
pump lever mechanism.
c. Push the polymer pouch lever up.
6. Tap Verify pouch to update the pouch information.
7. Tap Next to continue.
Reactivate Instrument—Install ABC
In this procedure, you install the anode buffer container (ABC).
IMPORTANT! If an autosampler error or a "power save" message is displayed, or if the CBC is not in
the forward, home position when you need to install the CBC, see "Check the position of the cathode
buffer container (CBC)"
1. Open the instrument doors.
2. Remove the waste ABC.
3. Re-install the ABC that was removed earlier, or install a new ABC. See "Install the anode buffer
container (ABC)"
1
Figure 86 Location of ABC and CBC
Anode buffer container (ABC)
1
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
on page
483.
on page
364.
Chapter 13 Maintain the Instrument
Maintenance wizard procedures
2
Cathode buffer container (CBC)
2
13
437

Advertisement

Table of Contents
loading

This manual is also suitable for:

A50369

Table of Contents