Cloud Analysis Troubleshooting; Consumables Troubleshooting - Thermo Scientific appliedbiosystems SeqStudio Flex Series User Manual

Genetic analyzer with instrument software v1.0
Table of Contents

Advertisement

Cloud analysis troubleshooting

See:
• "Plate file troubleshooting"
• "Remote Monitoring software troubleshooting"
• "Thermo Fisher

Consumables troubleshooting

Observation
A message is displayed
that indicates that insufficient
polymer volume is available
when you start a run, even
though the injection limit and
the sample limit have not been
met
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
on page 501
Connect troubleshooting"
Possible cause
The Remove Bubbles
maintenance wizard was used
excessively.
Note: A polymer pouch
includes a small reserve
volume that is used for the
Remove Bubbles maintenance
wizard, which consumes ~350
μL of polymer. The reserve
volume is sufficient to run the
wizard ~4 times (including the
remove bubbles step during
other maintenance wizards). If
you manually run the Remove
Bubbles maintenance wizard
>4 times, the volume of
polymer that is available for
samples may be depleted.
on page 505
on page 524
Recommended action
Replace the polymer pouch. In the future,
use the Remove Bubbles maintenance wizard
sparingly.
Appendix A Troubleshooting

Cloud analysis troubleshooting

A
495

Advertisement

Table of Contents
loading

This manual is also suitable for:

A50369

Table of Contents