Thermo Scientific appliedbiosystems SeqStudio Flex Series User Manual page 389

Genetic analyzer with instrument software v1.0
Table of Contents

Advertisement

2. Obtain the capillary array to install. Carefully remove the covers from the capillary array tip,
detection cell window, and array header. Retain the covers for future use.
Ensure that the capillary tips are not crushed or damaged.
1
Figure 33 Capillary array covers
Array tip cover
1
Detection cell cover (on the back of the assembly)
2
Array electrode cover with capillary tips submerged in distilled water
3
3. Open the array header latch by pulling the snap-pin connector.
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
2
Chapter 13 Maintain the Instrument
Maintenance wizard procedures
3
13
389