Chapter 13 Maintain the Instrument
13
Clean the instrument
d. Clean the CBC holder with a moistened lint-free lab wipe, wipe off any liquid on the CBC, then
replace the CBC.
IMPORTANT! To avoid injury, do not place your hands into any empty spaces in the drawer.
If you observe liquid on any surface that you cannot wipe, contact Technical Support.
4. Clean off any polymer crystals on the instrument, including the capillary tips, with deionized water
and lint-free tissue.
5. Clean the array port lock with deionized water and lint-free tissue. See Figure 18
6. Clean the drip tray with deionized water and a lint-free tissue.
Note: The drip tray can be removed. See Figure 18
Figure 18 Array port lock and drip tray location
Array port lock
1
362
1
2
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
on page
362.
Drip tray
2
on page
362.